Choosing a dose-setting method: VDmax25 or Method 1
How the two most widely used ISO 11137-2 methods work, what each costs in samples and time, and how to choose between them.
Before a medical device can be sterilized with gamma radiation, ISO 11137 requires the sterilization dose to be established and then maintained. The two most widely used dose-setting approaches in ISO 11137-2 are VDmax25 and Method 1. Choosing the right one saves time, samples and money.
VDmax25: substantiating 25 kGy
VDmax25 confirms that 25 kGy achieves a sterility assurance level (SAL) of 10−6 for your product. It suits products with an average bioburden of up to 1,000 CFU.
- Bioburden is determined on items from three production batches.
- A verification dose, read from the tables in ISO 11137-2 according to the bioburden, is applied to 10 product items.
- The test passes if no more than 1 of the 10 items is positive in the test of sterility.
Strengths: few samples, and a fast route to a validated dose. Limitation: it commits you to a minimum dose of 25 kGy, which can matter for radiation-sensitive materials.
Method 1: a product-specific dose
Method 1 uses your product's bioburden, together with a standard distribution of microbial resistance, to set a verification dose at SAL 10−2. It then establishes the sterilization dose for SAL 10−6.
- Bioburden is determined on items from three production batches.
- The verification dose is applied to 100 product items, and is accepted if no more than 2 are positive.
- For low-bioburden products, the resulting dose can be well below 25 kGy.
Strengths: lower doses protect sensitive polymers and can reduce processing cost at volume. Limitation: 100 samples, with more test cost and time.
Side by side
| VDmax25 | Method 1 | |
|---|---|---|
| Items at verification dose | 10 | 100 |
| Acceptance | ≤ 1 positive | ≤ 2 positives |
| Bioburden range | Average up to 1,000 CFU | Covered by the method's tables |
| Sterilization dose | 25 kGy minimum | Product-specific, can be lower |
| Best for | Robust materials, fast validation | Radiation-sensitive materials, high volumes |
After validation: dose audits
The established dose has to be maintained through periodic sterilization dose audits. ISO 11137-1 sets the starting audit interval at no more than three months; the frequency can be reduced only when it is justified by data, as described in the standard. A failed audit triggers an investigation and, where needed, dose augmentation.
Related reading.
Dose mapping explained: Dmin, Dmax and the dose uniformity ratio
Why no carton receives a perfectly even dose, how dose mapping finds the minimum and maximum, and what the dose uniformity ratio means for your product.
Packing cartons for gamma: seven rules for consistent dose
Practical packing rules that keep your product inside its validated dose window, batch after batch.
Gamma or ethylene oxide: which is right for your product?
The two most common industrial sterilization methods compared on penetration, residues, release time, packaging and material compatibility.