Choosing a dose-setting method: VDmax25 or Method 1

How the two most widely used ISO 11137-2 methods work, what each costs in samples and time, and how to choose between them.

Gammabrau Technical Team2 min read LinkedInWhatsApp

Before a medical device can be sterilized with gamma radiation, ISO 11137 requires the sterilization dose to be established and then maintained. The two most widely used dose-setting approaches in ISO 11137-2 are VDmax25 and Method 1. Choosing the right one saves time, samples and money.

VDmax25: substantiating 25 kGy

VDmax25 confirms that 25 kGy achieves a sterility assurance level (SAL) of 10−6 for your product. It suits products with an average bioburden of up to 1,000 CFU.

  • Bioburden is determined on items from three production batches.
  • A verification dose, read from the tables in ISO 11137-2 according to the bioburden, is applied to 10 product items.
  • The test passes if no more than 1 of the 10 items is positive in the test of sterility.

Strengths: few samples, and a fast route to a validated dose. Limitation: it commits you to a minimum dose of 25 kGy, which can matter for radiation-sensitive materials.

Method 1: a product-specific dose

Method 1 uses your product's bioburden, together with a standard distribution of microbial resistance, to set a verification dose at SAL 10−2. It then establishes the sterilization dose for SAL 10−6.

  • Bioburden is determined on items from three production batches.
  • The verification dose is applied to 100 product items, and is accepted if no more than 2 are positive.
  • For low-bioburden products, the resulting dose can be well below 25 kGy.

Strengths: lower doses protect sensitive polymers and can reduce processing cost at volume. Limitation: 100 samples, with more test cost and time.

Side by side

VDmax25Method 1
Items at verification dose10100
Acceptance≤ 1 positive≤ 2 positives
Bioburden rangeAverage up to 1,000 CFUCovered by the method's tables
Sterilization dose25 kGy minimumProduct-specific, can be lower
Best forRobust materials, fast validationRadiation-sensitive materials, high volumes

After validation: dose audits

The established dose has to be maintained through periodic sterilization dose audits. ISO 11137-1 sets the starting audit interval at no more than three months; the frequency can be reduced only when it is justified by data, as described in the standard. A failed audit triggers an investigation and, where needed, dose augmentation.

How we helpGammabrau supports dose setting, dose mapping and audit irradiations, coordinated with your microbiology testing laboratory, and supplies the dosimetry records your technical file needs.