Sterile to SAL 10−6, in the sealed box.
Gamma rays from Cobalt-60 pass straight through cartons, pouches and blister packs, so devices are sterilized exactly as they ship: no unpacking, no heat, no gas and no residue. Each batch is released on measured dose, with a Certificate of Irradiation for your batch record.
Products we process.
Injection and infusion
Syringes, needles, IV and infusion sets, blood collection tubes and transfusion sets.
Surgical and wound care
Sutures, gauze, dressings, drapes, gowns and surgical gloves.
Catheters and tubing
Urinary and suction catheters, extension lines, connectors and tubing sets.
Laboratory and diagnostics
Petri dishes, pipettes, specimen containers and sample collection kits.
Procedure kits
Convenience and procedure packs with several components in one sealed pack.
Disposables
Masks, caps, shoe covers and other single-use protective items.
From first sample to routine batches.
Material compatibility
Your device and packaging are irradiated at the maximum dose they could receive, then tested for function, strength, colour and seal integrity.
Dose setting and mapping
The sterilization dose is established by VDmax25 or Method 1 to ISO 11137-2, and a dose map in your loading pattern proves every unit falls inside the window.
Routine processing
Batches are processed to the validated specification, released on routine dosimetry, and the dose is maintained through periodic dose audits.
Measured, certified, close by.
Licensed and certified
CDSCO licence MFG/MD/2022/000246 (Class C, Form MD-9) and BSI-certified ISO 13485:2016 (MD 737411).
Validation support
Dose setting, dose mapping and audit irradiations, coordinated with your microbiology testing laboratory.
Central India location
Betma, Indore, close to the Pithampur industrial area and on the main routes to Gujarat and Maharashtra.
Common questions.
What dose is used to sterilize medical devices?
Many devices are sterilized at 25 kGy, substantiated by the VDmax25 method in ISO 11137-2. Low-bioburden or radiation-sensitive products can be validated at a lower, product-specific dose using Method 1.
Will gamma damage my device materials?
Most medical plastics, including polyethylene, polystyrene, polycarbonate and radiation-grade polypropylene, tolerate sterilization doses well. PTFE and some unstabilised polypropylene grades do not. Test at the maximum dose before you commit.
What documents will I receive?
A Certificate of Irradiation for every batch, showing the dose required and the minimum and maximum dose delivered. Dosimetry records and our facility certificates are available for your technical file and audits.
Can you help with ISO 11137 validation?
Yes. We support dose setting, dose mapping and quarterly dose audits, and work with your QA team and testing laboratory on bioburden and sterility testing.
From our Insights.
Choosing a dose-setting method: VDmax25 or Method 1
How the two most widely used ISO 11137-2 methods work, what each costs in samples and time, and how to choose between them.
Dose mapping explained: Dmin, Dmax and the dose uniformity ratio
Why no carton receives a perfectly even dose, how dose mapping finds the minimum and maximum, and what the dose uniformity ratio means for your product.
Gamma or ethylene oxide: which is right for your product?
The two most common industrial sterilization methods compared on penetration, residues, release time, packaging and material compatibility.